Flare
Kurz Instruments, Inc.
Application Summary
MFT / 454FT Application Summary
Kurz Instruments, Inc. 454FT Insertion
Flow Meters used in Flare Applications
A chemical plant in Louisiana wanted a better handle
on Flare events. The Plant already had flow meters on the
main flare headers but during flare events, someone had
to “search” for the area and/or vessel that was causing
the event.
After some consideration as to the measurements accuracy
and cost requirement, Kurz 454FT insertion flow meters were
purchased and installed on each line feeding the main flare
headers.
These meters are used for both accurate identification
of the area causing the event and to provide a flow signal/totalization
of the flared gas.
All of the meters are calibrated for Methane and a mole
weight correction is used to correct the flow signal to
the proper gas being flared.
Mole weight correction is accomplished by using a multiplier
based calculating the square root of the mole weight of
the measured gas divided by methane’s mole weight.
It should be noted that some gases, such as hydrogen,
have very different thermal properties and will cause an
error in the above calculation.
Application Specifics:
Line sizes from 3-18”
Methane Correlation calibrations
Packing Glands and Retractor Restraints are used
Customer Benefits:
Faster response to flare events.
Measurement is used for reporting purposes.
|